Cirugía de microstent supraciliar ab interno para el glaucoma de ángulo abierto

Contraer todo Desplegar todo

Antecedentes

El glaucoma es la principal causa de ceguera global irreversible, a menudo asociado con una presión intraocular (PIO) elevada. Cuando el tratamiento médico o con láser ha fracasado o no se tolera, a menudo es necesaria la cirugía. En los últimos años se han desarrollado métodos quirúrgicos mínimamente invasivos para reducir la PIO con menores riesgos quirúrgicos. La cirugía de microstent supraciliar para el tratamiento del glaucoma de ángulo abierto (GAA) es uno de estos métodos.

Objetivos

Evaluar la eficacia y seguridad de la cirugía de microstent supraciliar para el tratamiento del GAA, y compararla con los tratamientos médicos, con láser o quirúrgicos estándar.

Métodos de búsqueda

Se hicieron búsquedas en el Registro Cochrane central de ensayos controlados (Cochrane Central Register of Controlled Trials, CENTRAL; que contiene el Registro de ensayos del Grupo Cochrane de Salud ocular y de la visión [Cochrane Eyes and Vision]; 2020, número 8); Ovid MEDLINE; Ovid Embase; el registro ISRCTN; ClinicalTrials.gov y la ICTRP de la OMS. La fecha de la búsqueda fue el 27 de agosto de 2020.

Criterios de selección

Se buscaron ensayos controlados aleatorizados (ECA) de cirugía de microstent supraciliar, sola o con cirugía de cataratas, en comparación con otros tratamientos quirúrgicos (cirugía de cataratas sola, otras técnicas mínimamente invasivas con dispositivos para el glaucoma, trabeculectomía), tratamiento con láser o tratamiento médico.

Obtención y análisis de los datos

Dos autores de la revisión de forma independiente examinaron los títulos y resúmenes de la búsqueda en la base de datos para identificar los estudios que cumplían los criterios de selección. La extracción de los datos, el análisis y la evaluación del riesgo de sesgo de los estudios seleccionados se realizaron de forma independiente y según la metodología Cochrane estándar.

Resultados principales

Un estudio cumplió con los criterios de inclusión de esta revisión, y evaluó la eficacia y la seguridad de la cirugía con microstent supraciliar Cypass para el tratamiento del GAA, comparando la facoemulsificación + la cirugía con microstent supraciliar con la facoemulsificación sola durante 24 meses. Este estudio incluyó 505 ojos de 505 participantes con GAA y cataratas, 374 asignados al azar al grupo de facoemulsificación + microstent.

En este estudio, el riesgo de sesgo percibido a partir de la generación de la secuencia aleatoria, la ocultación de la asignación y el informe selectivo fue bajo. Sin embargo, se consideró que el estudio tenía un alto riesgo de sesgo de realización, ya que los cirujanos/investigadores no estaban cegados. El sesgo de desgaste fue incierto; 448/505 participantes contribuyeron al análisis por protocolo.

La inserción de un microstent supraciliar Cypass combinada con la facoemulsificación probablemente aumenta la proporción de participantes que no se tienen que administrar medicación (no utilizan gotas para los ojos) a los 24 meses en comparación con la facoemulsificación sola (85% versus 59%, razón de riesgos [RR]1,27; intervalo de confianza [IC] del 95%: 1,09 a 1,49, evidencia de certeza moderada).

Hay evidencia de certeza alta de que se produce una mayor mejoría de la PIO media en el grupo de facoemulsificación + microstent: cambio medio (DE) de la PIO desde el inicio de ‐5,4 (3,9) mmHg en el grupo de facoemulsificación, en comparación con ‐7,4 (4,4) mmHg en el grupo de facoemulsificación + microstent a los 24 meses (diferencia de medias ‐2,0 mmHg; IC del 95%: ‐2,85 a ‐1,15).

Hay evidencia de certeza moderada de que la inserción de un microstent se asocia probablemente con una mayor reducción del uso de gotas para reducir la PIO (reducción media de 0,7 medicamentos en el grupo de facoemulsificación, en comparación con una reducción media de 1,2 medicamentos en el grupo de facoemulsificación + microstent).

La inserción de un microstent durante la facoemulsificación podría reducir la necesidad de una intervención posterior para controlar la PIO en comparación con la facoemulsificación sola (RR 0,26; IC del 95%: 0,07 a 1,04; evidencia de certeza baja).

No hay evidencia relacionada con la tasa de progresión del campo visual, o la proporción de participantes cuya pérdida del campo visual progresó en este estudio.

Hay evidencia de certeza moderada que muestra poca o ninguna diferencia en la proporción de participantes que presentan complicaciones posoperatorias en el transcurso de 24 meses entre los participantes del grupo de microstent en comparación con los que solo recibieron facoemulsificación (RR 1,1; IC del 95%: 0,8 a 1,4).

Los datos posteriores a la aprobación de cinco años relacionados con la seguridad del microstent supraciliar Cypass mostraron un aumento de la pérdida de células endoteliales, asociado con la posición del microstent en la cámara anterior.

En el estudio incluido no se informaron resultados de la calidad de vida relacionada con la salud (CdVRS).

Conclusiones de los autores

Los datos de este único ECA muestran la superioridad de la cirugía de microstent supraciliar cuando se combina con la facoemulsificación, en comparación con la facoemulsificación sola para lograr el control sin medicación del GAA. Sin embargo, el dispositivo utilizado en este ensayo plantea problemas de seguridad a largo plazo, relacionados con una pérdida significativa de células endoteliales de la córnea, observada a los cinco años de la implantación del dispositivo. En el momento de realizar esta revisión, este dispositivo se ha retirado del mercado.

Esta revisión ha encontrado que existen pocos estudios de calidad alta que comparen la cirugía de microstent supraciliar con los tratamientos médicos, con láser o quirúrgicos estándar para el glaucoma. Este hecho se debería abordar mediante la realización de más ECA adecuadamente diseñados con un seguimiento a largo plazo suficiente para garantizar la obtención de datos de seguridad sólidos. La consideración de los resultados de la calidad de vida relacionada con la salud también debería figurar en el diseño de los ensayos.

PICO

Population
Intervention
Comparison
Outcome

El uso y la enseñanza del modelo PICO están muy extendidos en el ámbito de la atención sanitaria basada en la evidencia para formular preguntas y estrategias de búsqueda y para caracterizar estudios o metanálisis clínicos. PICO son las siglas en inglés de cuatro posibles componentes de una pregunta de investigación: paciente, población o problema; intervención; comparación; desenlace (outcome).

Para saber más sobre el uso del modelo PICO, puede consultar el Manual Cochrane.

Resumen en términos sencillos

¿La colocación de un tubo diminuto (microstent) bajo la superficie del ojo alivia la presión elevada de larga duración dentro del ojo (glaucoma)?

¿Qué es el glaucoma de ángulo abierto?
El glaucoma es una enfermedad ocular frecuente causada por la acumulación de líquido en la parte frontal del ojo, lo que aumenta la presión dentro del ojo. El aumento de la presión daña el nervio que conecta el ojo con el cerebro (nervio óptico) y provoca la pérdida de visión. El glaucoma puede provocar la pérdida permanente de la visión (ceguera) si no se diagnostica y trata a tiempo.

El glaucoma de ángulo abierto es el tipo más común de glaucoma y se tiende a desarrollar lentamente durante muchos años. La causa es que los canales de drenaje del ojo se obstruyen gradualmente con el tiempo.

Tratamientos para el glaucoma
El tratamiento no puede revertir la pérdida de la visión que se produjo antes de que se diagnosticara el glaucoma, pero puede enlentecer o detener la pérdida de la visión. Todos los tratamientos del glaucoma tienen como objetivo reducir la presión del ojo. Estos incluyen:

‐ medicamentos, administrados en forma de gotas para los ojos;
‐ tratamiento con láser para reducir la producción de líquido y abrir los canales de drenaje bloqueados; o
‐ cirugía para drenar el líquido del ojo.

Uno de los tratamientos consiste en colocar un pequeño tubo (llamado microstent) bajo la superficie del ojo para crear un canal de drenaje para el exceso de líquido. Los microstents se suelen colocar durante una intervención quirúrgica para tratar las cataratas, es decir, las manchas nubladas que se forman en el interior del cristalino y que hacen que la vista se vuelva borrosa.

¿Por qué se ha elaborado esta revisión Cochrane?
La colocación de un microstent podría reducir la presión en el interior del ojo y disminuir la necesidad de medicamentos en forma de gotas o de otros tipos de cirugía que podrían tener mayores riesgos. Se quería determinar si la colocación de un microstent durante la cirugía de cataratas reduciría la presión dentro del ojo en personas con glaucoma de ángulo abierto.

También interesaba saber cómo el microstent afectaba:

‐ la necesidad de medicamentos para reducir la presión en el ojo; y
‐ el bienestar de las personas.

¿Qué se hizo?
Se buscaron estudios que probaran el efecto de colocar un microstent durante la cirugía de cataratas en personas con glaucoma de ángulo abierto. Se buscaron estudios controlados aleatorizados en los que las personas que recibieron un microstent y las que no lo recibieron se decidieron por azar. Este tipo de estudios suele proporcionar la evidencia más fiable sobre los efectos de un tratamiento.

Fecha de búsqueda: se incluyó la evidencia publicada hasta agosto de 2020.

Qué se encontró
Se encontró un estudio realizado en EE.UU. en el que participaron 505 personas (de 45 años o más) con glaucoma de ángulo abierto y cataratas.

El estudio dividió a los pacientes en dos grupos. A un grupo se le colocó un microstent durante la cirugía para tratar la catarata; el otro grupo fue operado solo para tratar la catarata. Los pacientes del estudio se evaluaron durante dos años.

El estudio fue financiado por una empresa que fabrica microstents para el tratamiento del glaucoma.

¿Cuáles son los resultados principales de esta revisión?
Dos años después de someterse a una operación de cataratas, en personas a las que también se les colocó un microstent:

‐ más de ellos (el 85% en este grupo en comparación con el 59% en el otro) no necesitaron medicamentos en forma de gotas para tratar el glaucoma (evidencia de 448 personas);
‐ tuvieron mayores reducciones de la presión dentro del ojo afectado, que las personas a las que no se les colocó un microstent (448 personas);
‐ tuvieron una mayor reducción, como promedio, en el uso de medicamentos en forma de gotas para los ojos, que las personas a las que no se les colocó un microstent (448 personas); y
‐ menos personas necesitaron una nueva cirugía para tratar el glaucoma (505 personas).

Sin embargo, la colocación del microstent provocó un mayor número de efectos no deseados (complicaciones) en los dos años posteriores a la cirugía, en comparación con la cirugía de cataratas sola (evidencia de 505 personas). Como promedio, por cada 1000 personas, 390 personas a las que se les aplicó el microstent tuvieron complicaciones, en comparación con 360 personas a las que no se les aplicó el microstent. Existe la preocupación de que el microstent utilizado en este estudio cause daños duraderos en la capa transparente de la parte delantera del ojo (córnea).

El estudio no midió el bienestar de las personas (calidad de vida) ni midió cómo se vio afectada la vista de las personas durante los dos años posteriores a la cirugía.

Confianza en estos resultados
Existe seguridad con respecto a las reducciones de la presión en el interior del ojo y las complicaciones después de la cirugía. No se espera que nueva evidencia cambie estos resultados.

Se tiene una confianza moderada en las reducciones de la necesidad de medicamentos en forma de gotas para reducir la presión dentro del ojo. Aunque los pacientes del estudio no sabían en qué grupo de tratamiento se encontraban, las personas que administraban los tratamientos sí lo sabían y esto podría haber afectado los resultados. Estos resultados podrían cambiar si se dispusiera de más evidencia.

Existe menos seguridad con respecto a cuántas personas necesitaron cirugía adicional para tratar el glaucoma y es probable que más evidencia cambie estos resultados.

Mensajes clave
La colocación de un microstent en el ojo durante la cirugía de cataratas reduce la presión dentro del ojo en personas con glaucoma de ángulo abierto, y reduce la necesidad de medicamentos para reducir la presión, más que la cirugía de cataratas sola. No obstante, la colocación del microstent se relacionó con un aumento de las complicaciones tras la cirugía.

Authors' conclusions

Implications for practice

This review has identified RCT data showing a superiority in effectiveness of supraciliary microstent surgery when combined with phacoemulsification compared to phacoemulsification alone in achieving drop‐free control of OAG, an aspect of importance to people with OAG. However, there are associated safety concerns with the device used in the single published RCT (COMPASS Trial), with particular focus on the health of the cornea endothelium after device implantation leading to its withdrawal form the market at the time of this review.

Additionally, this review highlights the lack of high‐quality trial data comparing supraciliary microstent surgery to standard medical, laser or surgical glaucoma treatments. This should be a consideration when clinicians and other decision makers discuss management options with people in the treatment of OAG.

Implications for research

This review demonstrates that RCTs can be performed to assess the effectiveness and safety of supraciliary microstents, one of several minimally‐invasive glaucoma devices proposed as alternatives to standard glaucoma interventions. However, as this review highlights, there is a lack of high‐quality trial data comparing supraciliary microstent surgery to standard medical, laser or surgical glaucoma treatments that should be addressed.

The emergence of safety concerns in the five‐year post‐approval safety study for the Cypass supraciliary microstent device also reminds us of the important value of long‐term study data, particularly in the assessment of new interventions.

Although clinical outcome measures are important, future study design should also consider including outcome measures on health‐related quality of life and other aspects important to people with OAG.

Summary of findings

Open in table viewer
Summary of findings 1. Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone for open‐angle glaucoma, at 24 months

Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone for open‐angle glaucoma

Patient or population: people with open‐angle glaucoma
Setting: hospital or outpatient clinic
Intervention: phacoemulsification + supraciliary microstent surgery
Comparison: phacoemulsification alone

Outcomes

Anticipated absolute effects* (95% CI)

Relative effect
(95% CI)

№ of participants
(studies)

Certainty of the evidence
(GRADE)

Comments

Risk with phacoemulsification alone

Risk with phacoemulsification + supraciliary microstent surgery

Proportion of participants who were medication‐free (not using eye‐drops)
at 24 months (medium‐term)

Study population

RR 1.27
(1.09 to 1.49)

448
(1 RCT)

⊕⊕⊕⊝
MODERATE1

 

595 per 1000 (500 to 685)

849 per 1000 (806 to 886)

Mean change in unmedicated IOP
24 months (medium‐term)

Study population

MD ‐2.0 mmHg  (‐2.85 to ‐1.15)

448
(1 RCT)

⊕⊕⊕⊕
HIGH2

 

The mean change (reduction) in IOP in the control group at 24 months was 5.4 (SD 3.9) mmHg

The mean change (reduction) in IOP in the intervention group at 24 months was 7.4 (SD 4.4) mmHg

Mean change in daily IOP‐lowering medications
at 24 months (medium‐term)

Study population

MD ‐0.5 medications (‐0.68 to ‐0.32)

448
(1 RCT)

⊕⊕⊕⊝
MODERATE1

 

The mean reduction in number of IOP‐lowering drops was 0.7 medications

The mean reduction in number of IOP‐lowering drops was 1.2 medications

Proportion of participants who required further glaucoma surgery
at 24 months (medium‐term)

Study population

RR 0.26 (0.07 to 1.04)

505
(1 RCT)

⊕⊕⊝⊝
LOW3

 

31 per 1,000

8 per 1,000

(2 to 32)

 

 

Mean change in health‐related quality of life

The included study did not report this outcome.

Rate of visual field progression or proportion of participants whose field loss progressed 

The included study did not report this outcome.
 

Proportion of participants experiencing postoperative complications
over 24 months (medium‐term)

Study population

RR 1.1 (0.8 to 1.4)

505
(1 RCT)

⊕⊕⊕⊕
MODERATE4

Five year post‐approval data regarding the safety of the Cypass supraciliary microstent showed increased endothelial cell loss, associated with the position of the microstent in the anterior chamber.

360 per 1,000

390 per 1,000

*The risk in the intervention group (and its 95% confidence interval) is based on the assumed risk in the comparison group and the relative effect of the intervention (and its 95% CI).

CI: confidence interval; IOP: intraocular pressure; MD: mean differenceOR: odds ratio; RR: risk ratio; SD: standard deviation

GRADE Working Group grades of evidence
High‐certainty: we are very confident that the true effect lies close to that of the estimate of the effect.
Moderate‐certainty: we are moderately confident in the effect estimate: The true effect is likely to be close to the estimate of the effect, but there is a possibility that it is substantially different.
Low‐certainty: our confidence in the effect estimate is limited: The true effect may be substantially different from the estimate of the effect.
Very low‐certainty: we have very little confidence in the effect estimate: The true effect is likely to be substantially different from the estimate of effect.

1Downgraded one level for study limitations: although participants were masked to their treatment group throughout the study period, as were IOP reading technicians, surgeons/Investigators were not.

2Not downgraded for study limitations as IOP assessment was masked.

3Downgraded two levels: one for imprecision ‐ confidence intervals included 1, no effect, and one level for risk of bias.

4Downgraded one level for imprecision: confidence intervals included 1, no effect.

Background

Description of the condition

Glaucoma is a chronic progressive optic neuropathy, affecting up to 4% of people by the age of 80 years (Burr 2007). It is the leading cause of irreversible blindness, affecting 60 million people globally (Quigley 2006). This figure is expected to increase to 80 million people by 2020. Open‐angle glaucoma (OAG) is the commonest type, accounting for three‐quarters of cases (Quigley 2006). In one large population cohort, one in six patients with OAG became bilaterally blind (Peters 2013). The only proven way to prevent vision loss is to reduce the pressure inside the eye (intraocular pressure) over the long term (AGIS 2000; CNTG Study Group 1998; Heijl 2002; Kass 2002). Approaches to reducing intraocular pressure (IOP) include medical therapy, laser treatments, and surgery. Commercially available eye‐drop preparations have a short‐lasting effect; medical therapy requires eye‐drops to be instilled one or more times daily for life. Adherence is very poor, even if use is monitored (Friedman 2009; Okeke 2009). Conventional surgical techniques such as trabeculectomy are associated with significant risks, with more than 40% of patients developing perioperative complications (Kirwan 2013; Lichter 2001) and re‐operation being needed in 7% to 18% (Gedde 2012; Kirwan 2013). Therefore, they are often reserved for disease that is progressing despite other treatments (King 2013).

Description of the intervention

Recently, a number of minimally‐invasive surgical techniques have been developed with the aim of achieving long‐term reduction of IOP with a better safety profile than conventional surgery (Francis 2011). Among them is ab interno supraciliary microstent surgery ‐ the Cypass Microstent (Alcon Laboratories, a division of Novartis, Basel, Switzerland) and the iStent Supra (Glaukos Corporation, Laguna Hills, CA, USA) are examples of these devices. The former is FDA approved and also CE (European Conformity) marked in Europe. The latter is undergoing a phase 3 clinical trial with a view to obtaining FDA approval, but is CE marked in Europe.

How the intervention might work

In cases of open‐angle glaucoma, an increased resistance to outflow is thought to exist not only at the level of the trabecular meshwork but also within the ciliary body part of the uveoscleral pathway.

With the uveoscleral pathway thought to contribute up to half of physiological aqueous outflow (Toris 1999), supraciliary microstents such as the Cypass and iStent Supra have been developed to bypass this, leading to an increase in aqueous outflow and a reduction in intraocular pressure.

Why it is important to do this review

Consultation with patients and healthcare professionals has identified a need for better treatments for glaucoma (James Lind Alliance 2013). Minimally‐invasive glaucoma procedures allow the possibility of safe and effective long‐term reduction of IOP, removing concerns about permanent vision loss due to non‐adherence to eye‐drops. A single treatment may also be more acceptable to patients than lifelong daily administration of eye‐drops.

The evidence base intended to support the use of supraciliary microstents in practice continues to grow. Randomised controlled clinical studies to assess the safety and efficacy of the Cypass and iStent Supra alone have recruited in excess of 1000 participants. However, what is less clear is where this evidence lies in the current landscape of existing interventional options to manage open‐angle glaucoma, presently including medical, laser, trabeculectomy and other minimally‐invasive glaucoma procedures. Since phacoemulsification alone has been shown to  reduce IOP (Mansberger 2012), we specifically examined the evidence for the efficacy of supraciliary drainage devices when combined with phacoemulsification in comparison to phacoemulsification alone.

With both the Cypass and iStent Supra devices holding a CE mark for use in Europe and the Cypass already FDA approved, the user availability of such supraciliary microstents is expected to grow in the coming years, increasing the importance of a review that will critically evaluate the current evidence relating to this group of devices.

This Cochrane review was conducted in parallel with other reviews currently undertaken by the Cochrane Eyes and Vision MIGS Consortium, which includes minimally‐invasive glaucoma surgery (MIGS) techniques and devices such as the Trabectome (NeoMedix, Tustin, California) (Hu 2021), Hydrus Schlemm's canal Microstent (Ivantis Inc., Irvine, California) (Otarola 2020), endoscopic cytophotocoagulation (ECP) (Endo Optiks, Waltham, Massachusetts) (Tóth 2019), XEN Glaucoma Implant (Allergan, Dublin, Ireland) (King 2018) and IStent or IStent inject (Glaukos Corporation, Laguna Hills, California) (Le 2019).

Objectives

To evaluate the efficacy and safety of supraciliary microstent surgery for the treatment of OAG, and to compare with standard medical, laser or surgical treatments.

Methods

Criteria for considering studies for this review

Types of studies

We included randomised controlled trials (RCTs) reported in any language irrespective of their publication status. 

Types of participants

Study participants had OAG of any type, including primary and secondary OAG. Closed angle glaucoma was excluded. As there are no universally‐accepted criteria by which glaucoma may be defined, we permitted studies to use their own definitions of glaucoma (provided these were clearly stated). In addition, participants with ocular hypertension, normal tension glaucoma, or possible glaucoma (suspects for glaucoma) were included. We did not apply any restrictions regarding location, setting, or demographic factors.

Types of interventions

We compared ab interno supraciliary microstent surgery with the Cypass (Alcon Laboratories, a division of Novartis, Basel, Switzerland), iStent Supra (Glaukos Corporation, Laguna Hills, CA, USA) or other supraciliary microstents that were identified during this review to:

  • laser treatment (selective laser trabeculoplasty or argon laser trabeculoplasty);

  • other minimally‐invasive glaucoma surgery (MIGS) techniques;

  • conventional glaucoma surgery (trabeculectomy);

  • medical therapy.

RCTs were considered where supraciliary microstent devices were used in combination with phacoemulsification, as well as RCTs where these devices were used in isolation. 

Types of outcome measures

We did not use the reporting of particular outcomes as a criterion for eligibility for review. We did not exclude studies from review solely on the grounds of an outcome of interest not being reported.

We planned to report outcomes in the short‐term (six to 18 months), medium‐term (18 to 36 months), and long‐term (36 months onwards).

Primary outcomes

  • Proportion of participants who were medication‐free (not using eye‐drops).

Several different glaucoma outcome measures have been specified as primary outcomes in other Cochrane Reviews and protocols (Ismail 2015). A recent study classified IOP, visual field, safety, and anatomic outcomes as being highly important to glaucoma experts (Ismail 2016). A panel of patients from the Patient and Public Involvement Group of the National Institute for Health Research (NIHR) Biomedical Research Centre for Ophthalmology identified drop‐free disease control as a highly valued outcome (unpublished). We chose a participant‐centred primary outcome.

Secondary outcomes

  • Mean change in IOP, measured using Goldmann applanation tonometry;

  • Mean change in number of IOP‐lowering drops taken per day;

  • Proportion of participants who achieved an IOP 21 mmHg or less;

  • Proportion of participants who achieved an IOP 17 mmHg or less;

  • Proportion of participants who achieved an IOP 14 mmHg or less;

  • Proportion of participants who required further glaucoma surgery, including laser, as recorded by the investigators of the included trial;

  • Rate of visual field progression (decibels (dB)/time) or proportion of participants whose field loss progressed in the follow‐up period;

  • Mean change in health‐related quality of life (HRQoL).

Adverse effects

  • Proportion of participants experiencing intraoperative and postoperative complications, including, but not restricted to, the following:

    • loss of visual acuity (more than 2 Snellen lines or more than 0.3 logMAR, according to the method of recording visual acuity; or loss of light perception);

    • bleeding, as recorded by the investigators;

    • endophthalmitis, as recorded by the investigators;

    • IOP spikes (postoperative rise in IOP, measured using Goldmann applanation tonometry, of more than 10 mmHg compared to the previous assessment, including measurements taken during the first postoperative month).

Search methods for identification of studies

Electronic searches

The Cochrane Eyes and Vision Information Specialist searched the following databases for randomised controlled trials and controlled clinical trials. There were no restrictions to language or year of publication. The date of the search was 27 August 2020.

  • Cochrane Central Register of Controlled Trials (CENTRAL; 2020, Issue 8) (which contains the Cochrane Eyes and Vision Trials Register) in the Cochrane Library (searched 27 August 2020) (Appendix 1);

  • MEDLINE Ovid (1946 to 27 August 2020) (Appendix 2);

  • Embase Ovid (1980 to 27 August 2020) (Appendix 3);

  • ISRCTN registry (www.isrctn.com/editAdvancedSearch; searched 27 August 2020) (Appendix 4);

  • US National Institutes of Health Ongoing Trials Register ClinicalTrials.gov (www.clinicaltrials.gov; searched 27 August 2020) (Appendix 5);

  • World Health Organisation (WHO) International Clinical Trials Registry Platform (ICTRP) (www.who.int/ictrp; searched 27 August 2020) (Appendix 6).

Searching other resources

We searched the reference lists of included studies for other possible studies. We also searched the websites of the manufacturers of current ab interno supraciliary microstents (Alcon.com, Alcon Laboratories, a division of Novartis, Basel, Switzerland; Glaukos.com, Glaukos Corporation, Laguna Hills, CA, USA) for any information on forthcoming trials.

Data collection and analysis

Selection of studies

Two review authors working independently (AS, HJ) screened titles and abstracts of all articles identified by the search using web‐based online review management software (Covidence). If abstracts were not available, full‐text articles were screened. Two review authors (AS, HJ) independently assessed full‐text reports of all potentially eligible studies. If there was disagreement regarding eligibility, a third review author arbitrated. If any full‐text reports were rejected, the reasons for this were recorded in the Characteristics of excluded studies table.

Data extraction and management

We extracted data from reports of included studies using a data collection form. Two review authors (AS, HJ) worked independently to extract study characteristics from reports and entered the data into Review Manager 5 (RevMan 5) (Review Manager 2020). The same authors extracted the data for analyses, and one review author (AS) checked the data before entering it into Review Manager (RevMan 5). If there was disagreement, a third review author arbitrated.

The process included cross‐checking data entry independently using Covidence to support this. If there was disagreement, a third independent review author would arbitrate.

We presented the data collected in Appendix 7 in the Characteristics of included studies table. Where data on included studies were missing or unclear, we planned to contact the individuals or organisations involved to obtain clarification. We collected and used the most detailed numerical data available to facilitate analyses of included studies. We attempted to obtain these data from individuals or organisations in preference to less precise methods such as extracting numeric data from graphs. If this was necessary, two independent review authors extracted the data and a third review author arbitrated, in case of disagreement.

Assessment of risk of bias in included studies

We used the Cochrane 'Risk of bias' tool as described in Chapter 8 of the Cochrane Handbook for Systematic Reviews of Interventions (Higgins 2017) to assess the risk of bias and assign judgements of this for included studies. Two review authors (AS, HJ) performed this 'Risk of bias' assessment independently. In the event of a disagreement, a third review author was available to arbitrate.

Measures of treatment effect

The primary outcome was the proportion of participants who were medication‐free at the studies' end. We used a risk ratio as the treatment effect measure. In assessing this effect measure, we have reported how prescribing of IOP‐lowering eye‐drops was determined during follow‐up, where this information was available. We examined whether the people measuring IOP and those deciding upon the prescribing of IOP‐lowering eye‐drops were masked to treatment group.

We have also reported mean change in IOP from randomisation to the studies' end. Secondary safety outcomes were to be reported as risk ratios. Health‐related quality of life outcomes were to be reported as differences in means or risk ratios for continuous and binary data, respectively.

Unit of analysis issues

We assessed whether included studies had included one or two eyes from each participant and whether or not randomisation has been conducted at the level of the participant or the eye. There is a potential for medical treatments, such as topical beta blockers used for one eye, to influence the outcome in the other eye (Piltz 2000). Surgery to lower IOP in one eye may also affect the IOP of the fellow eye (Radcliffe 2010). Therefore, we have excluded studies that had adopted a paired eye design. In the event of a multiple arm study being identified, this could be included providing the respective study design was adequate to ensure independent analysis of each treatment group occurred.

Dealing with missing data

We endeavoured to minimise missing outcome data by contacting individuals and organisations to obtain them. If the data were unavailable but the level of missing data in each group and reasons for missing data in each group were similar, we simply analysed available‐case data if an intention‐to‐treat (ITT) analysis had not been performed. We reported if authors had conducted their own ITT analysis despite missing data, but intended to document whether they provided any justification for the method they had used to deal with missing data and whether they had compared their ITT result with an available‐case result.

Assessment of heterogeneity

We intended to assess the heterogeneity between trials by careful examination of the study reports, assessing forest plots and an examination of the I2 value, however, as only one RCT met the inclusion criteria of this review, this was not necessary.

Assessment of reporting biases

We planned to use a funnel plot to assess the risk of publication bias if there were more than 10 trials within our review.

Data synthesis

We planned to undertake a meta‐analysis where data appeared clinically, methodologically, and statistically homogeneous. We planned to check that participants, interventions, comparators, and outcomes were sufficiently similar to give a clinically meaningful result and that our I2 result did not indicate considerable inconsistency (i.e. I2 less than 50%). If all estimates were in the same direction, we would meta‐analyse even where heterogeneity was evident but would comment on this. We planned to use a random‐effects model unless there were fewer than three eligible studies, in which case, we would use a fixed‐effect model. As we found only one study, a fixed‐effects model was used.

Subgroup analysis and investigation of heterogeneity

No subgroup analyses were performed in this review.

Sensitivity analysis

We planned to assess the impact of including studies at high risk of bias for an outcome in one or more key domains. However, there were too few included studies to conduct such analyses. 

Summary of findings and assessment of the certainty of the evidence

We prepared a table to summarise the findings of the review, including the assessment of the certainty of evidence for all outcomes using the GRADE approach (GRADEpro).

We reported the following outcomes at medium‐term follow‐up (18 to 36 months) in the 'Summary of findings' table for each comparison listed in the Types of interventions:  Ab interno supraciliary microstent surgery compared with laser treatment, other MIGS techniques, conventional glaucoma surgery (trabeculectomy), or medical therapy.

  • Proportion of participants who were medication‐free (not using eye drops);

  • Mean change in IOP, measured using Goldmann applanation tonometry;

  • Mean change in number of IOP‐lowering drops taken per day;

  • Proportion of participants who required further glaucoma surgery, including laser;

  • Rate of visual field progression (decibels (dB)/time) or proportion of participants whose field loss progressed in the follow up period;

  • Mean change in health‐related quality of life;

  • Proportion of participants experiencing intraoperative and postoperative complications (any time point).

Results

Description of studies

Results of the search

The electronic searches identified 481 references (Figure 1). After 144 duplicates were removed, the Cochrane Information Specialist (CIS) screened the remaining 337 records and removed 157 references that were not relevant to the scope of the review. We screened the remaining 180 references and obtained five full‐text reports for further assessment. We identified four full‐text reports of one study that met the inclusion criteria (COMPASS Trial), which included additional safety extension reports of the same study (Reiss 2019Lass 2019). We identified one report of one ongoing study that potentially meets the inclusion criteria (NCT01461278). The findings of this study should be considered upon study completion (last trial update 3/2020).

Included studies

We included one RCT, the COMPASS Trial, comprising 505 eyes and participants. This prospective, randomised, multicentre, controlled, interventional study reported two‐year and, in a later publication, also five‐year safety and efficacy results. It was conducted across 24 sites in the USA. People aged 45 years or older with mild to moderate primary open‐angle glaucoma, baseline unmedicated IOP between 21 and 33 mmHg, and cataract (best corrected visual acuity of 6/12 or worse), were randomised to phacoemulsification only, or phacoemulsification combined with ab interno supraciliary microstent insertion.

The primary outcome was the percentage of participants achieving a ≥ 20% diurnal lowering of unmedicated IOP from baseline. Secondary outcomes included mean unmedicated change in IOP, percentage of eyes with unmedicated IOP ≥ 6 and ≤ 18 mmHg, and change in number of glaucoma medications required. Additionally, the incidence of ocular adverse events was also recorded, both at two years. See the Characteristics of included studies table for more information.

Ongoing studies

One ongoing study met our inclusion criteria but is yet to report its findings (NCT01461278). Information on this study was obtained from the clinicaltrials.gov registry and also the device company website. Recruitment into this phase three clinical trial has been completed (1200 participants). This is a prospective, randomised, single‐masked, controlled, parallel‐group, multicentre study to evaluate the safety and efficacy of the Glaukos Suprachoroidal stent model G3 (also known as the Istent Supra) in people with mild to moderate primary open‐angle glaucoma. Study completion is expected to be December 2020. See the Characteristics of ongoing studies for more information.

Excluded studies

We did not exclude any studies from this review.

Risk of bias in included studies

An assessment of the risk of bias for the included study (COMPASS Trial), is shown in Figure 2 and Figure 3.


Risk of bias graph: review authors' judgements about each risk of bias item presented as percentages across all included studies.

Risk of bias graph: review authors' judgements about each risk of bias item presented as percentages across all included studies.


Risk of bias summary: review authors' judgements about each risk of bias item for each included study.

Risk of bias summary: review authors' judgements about each risk of bias item for each included study.

Allocation

Low risk of bias ‐ after central randomisation, group assignment was informed only after completion of cataract surgery. Allocation concealment occurred up to this point.

The same randomisation was maintained throughout the five‐year COMPASS XT safety and effectiveness study extension.

Blinding

Performance bias

We considered the study to be at high risk of performance bias because, whilst participants and IOP reading technicians were masked to their treatment group throughout the study period, surgeons and investigators were not masked.

Detection bias

We considered the study to be at low risk for detection bias as the IOP reading technicians were masked to group assignment.

The COMPASS XT study extension results were unmasked observations.

Incomplete outcome data

In the COMPASS Trial, 88.7% of recruited study participants contributed to the per protocol analysis (448/505). We assessed attrition bias as unclear as details were not provided, although the rates of attrition in the groups were similar.

In the COMPASS XT study, only 282 of the 480 cases who completed the 24‐month COMPASS Trial agreed to enrol, 253 of these completing five years.

Selective reporting

We considered the study to be at low risk of bias as the results aligned accurately with the registered study design and stated outcome measures (NCT01085357, part of  COMPASS Trial).

Other potential sources of bias

Although an objective and structured algorithm was described for the reintroduction of IOP‐lowering medications, in participants with IOP > 18 and < 21 mmHg, decisions were made on a 'case by case basis' giving rise to potential bias given that decision‐making investigators may be aware of group assignment.

Effects of interventions

See: Summary of findings 1 Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone for open‐angle glaucoma, at 24 months

Phacoemulsification + supraciliary microstent versus phacoemulsification alone

Proportion of participants who are medication‐free (not using eye‐drops)

In the COMPASS Trial, of the 448 participants completing 24 months follow‐up per‐protocol, 59.1% of the control group (phacoemulsification alone) were drop‐free compared to 84.8% in the phacoemulsification combined with ab interno supraciliary microstent insertion group (risk ratio (RR) 1.27, 95% confidence interval (CI) 1.09 to 1.49) (Analysis 1.1) ‐ there was moderate‐certainty in this group of an effective intervention.

The 60‐month safety and effectiveness study did not state the percentage of participants remaining medication‐free to compare directly with the 24‐month data.

Mean change in unmedicated IOP measured using Goldmann applanation tonometry

In the COMPASS Trial, at 24 months, a mean (SD) change in unmedicated IOP from baseline of 5.4 (3.9) mmHg in the control group (n = 116) was reported, compared to 7.4 (4.4) mmHg in the phacoemulsification + microstent group (n = 332); mean difference ‐2.0 (95% CI ‐2.85 to ‐1.15) (Analysis 1.2).

At 60 months, descriptive analysis showed a mean medicated/unmedicated IOP reduction of 8.0 (95% CI 6.8 to 9.2) in the control (n = 52) and 8.4 (95% CI 7.8 to 8.9) in the microstent (n = 200) groups, respectively.

Mean change in number of IOP‐lowering medications taken per day

In the COMPASS Trial, at 24 months, mean (SD) IOP‐lowering medication use changed from 1.3 (1.0) medications at baseline to 0.6 (0.8)  in the control group (n = 116) ‐ a mean change of 0.7 medications, and in the phacoemulsification + microstent group (n = 332), IOP medication use changed from 1.4 (0.9) at baseline to 0.2 (0.6 ) at 24 months ‐ a mean change of 1.2 medications. This showed a mean difference of ‐0.50 medications (95% CI ‐0.68 to ‐0.32)(Analysis 1.3).

Proportion of participants who achieved an IOP ≤ 21 mmHg

This outcome was not reported in the included trial.

Proportion of participants who achieved an IOP ≤ 17 mmHg

This specific outcome measure was not reported in the included trial, however, 66.7% and 44.0% (n = 200; CI 37.0 to 51.2) of  the phacoemulsification + microstent group and 40.9% and 28.3% (n = 52; CI 16.8 to 42.3) of the control group achieved a medication free IOP ≤ 18 mmHg at 24 months and 60 months respectively.

Proportion of participants who achieved an IOP ≤ 14 mmHg

This outcome was not reported in the included trial.

Proportion of participants who required further glaucoma surgery, including laser, as recorded by the investigators of the included trial

In the COMPASS Trial, four participants from the control group (4/131) and three participants from the intervention group (3/374) required further intervention for IOP control in the intention‐to‐treat population (RR 0.26, 95% CI 0.07 to 1.04), although the nature of the intervention was not described.

Mean change in health‐related quality of life (HRQoL)

This outcome was not reported in the included trial.

Proportion of participants experiencing intraoperative and postoperative complications

See Table 1.

Open in table viewer
Table 1. Ocular adverse events after phacoemulsification + supraciliary microstent surgery versus phacoemulsification, at 60 months

 

Intervention   
 

Phacoemulsification surgery
 

Phacoemulsification surgery + microstent
 

%

n

%

n

Outcomes

 

 

 

 

BCVA loss of > 3 lines or more compared to best BCVA reported in COMPASS study

0

0

0.9

2

Retinal detachment

1.5

1

0

0

Treatment of elevated intraocular pressure not satisfactorily managed with ocular hypotensive medication

1.5

1

0.5

1

Macular oedema

1.5

1

1.4

3

Other maculopathies

1.5

1

1.4

3

Corneal oedema

0

0

1.4

3

Events requiring unplanned surgical intervention

1.5

1

0.9

2

Results from the COMPASS XT publication. Phacoemulsification n = 200; phacoemulsification + microstent group n = 53, completing 60 months

Unmasked observational data. Similar baseline characteristics between groups.

Approximately 20% of the 60‐month data from the 253 participants cases completing the COMPASS XT study was obtained retrospectively.

No participants in the control group and 1.1% of participants in the phacoemulsification + microstent group lost more than two lines of vision at 24 months. 

At 60 months, 6.0% of the control group and 11.2% of the phacoemulsification + microstent group lost more than two lines of vision. Two participants (0.9%) in the phacoemulsification + microstent group lost more than 3 lines of vision (epiretinal membrane; cystoid macular oedema).

No participants in the control group and 2.7% of participants in the phacoemulsification + microstent group developed hyphaema (described as transient intraoperative).

There were no reported cases of endophthalmitis in either assigned group.

Postoperative IOP spikes (IOP ≥ 10 mmHg above baseline) occurred transiently in 2.3% of participants in the control group and 4.3% of participants in the phacoemulsification + microstent group. Transient hypotony was reported in 2.9% of participants in the phacoemulsification + microstent group.

Seven (1.9%) (7/374) participants in the phacoemulsification + microstent group developed a cyclodialysis cleft, no associated hypotony occurred, and no additional surgical intervention was required.

Sixty‐month post‐surgery data from an FDA‐mandated post‐approval safety study (NCT03273907) identified an elevated rate of endothelial cell density (ECD) reduction, with 27.16% (44/162) of microstented cases showing > 30% loss (FDA 2018Reiss 2019Table 2). There appeared to be an association between the extent of protrusion of the microstent into the anterior chamber and the rate of ECD loss. Three participants showed asymptomatic evidence of focal cornea oedema in the region of the microstent.

Open in table viewer
Table 2. Corneal endothelial cell loss after phacoemulsification + supraciliary microstent or phacoemulsification alone

Mean endothelial cell density (cells/mm2)        

 

Baseline 

At 60 months 

mean

Lower CI

Upper CI

n

mean

Lower CI

Upper CI

n

Phacoemulsification

2434.5

2356.5

2512.4

67

2189.1

2069

2309.2

40

Phacoemulsification + supraciliary microstent

2432.6

2382.8

2482.4

214

1931.2

1851.2

2011.2

163

 

Change from baseline (cells/mm2)     

mean

Lower CI

Upper CI

n

Phacoemulsification

‐249.6

‐341

‐158.2

40

Phacoemulsification + supraciliary microstent

‐507.6

‐581.7

‐433.6

163

 

Proportion of eyes with > 30% reduction in endothelial cell density from baseline at 60 months    

%

n

%

n

%

n

Number of retention rings visible on gonioscopy

< 1 microstent retention ring

1 microstent retention ring

≥ 2 microstent retention rings

Phacoemulsification + supraciliary microstent

20.6

13/63

21.9

16/73

57.7

15/26

Phacoemulsification alone

10.0

4/40

Results from the COMPASS trial safety extension publication. Extended interval specular microscopy performed subsequent to enrolment into the COMPASS extension trial (COMPASS XT)

Unmasked observational data

Four (1.9%) cases required a microstent trimming procedure.

Discussion

Summary of main results

We found one completed RCT, the COMPASS Trial, evaluating the efficacy and safety of supraciliary microstent surgery for the treatment of OAG, comparing phacoemulsification + supraciliary microstent surgery with phacoemulsification alone.

This review found moderate‐certainty evidence that the insertion of a Cypass supraciliary microstent combined with phacoemulsification increased the proportion of participants who were medication‐free at medium‐term follow‐up from 59% to 85% (RR 1.27, 95% CI 1.09 to 1.49).

High‐certainty evidence shows that a greater improvement in mean IOP occurred in the phacoemulsification + microstent group ‐ mean (SD) change in IOP from baseline of ‐5.4 (3.9) mmHg in the control group, compared to ‐7.4 (4.4) mmHg in the phacoemulsification + microstent group at 24 months (mean difference ‐2.0, 95% CI ‐2.9 to ‐1.1).

Moderate‐certainty evidence shows that mean IOP‐lowering drop use in the phacoemulsification + microstent group was associated with a reduction of 1.2 medications compared to 0.7 medications in the control group.

Moderate‐certainty evidence indicates that fewer participants in the microstent group required further glaucoma intervention to control IOP at a later stage: three phacoemulsification + microstent participants (3/374) compared to four control participants (4/131).

There is moderate‐certainty evidence relating to the proportion of participants experiencing postoperative complications over 24 months (medium‐term): anticipated absolute effect (95% CI) of 360 per 1000 and 390 per 1000 in the control and phacoemulsification plus microstent groups, respectively.

Concerns have emerged from five‐year post‐approval data regarding the safety of the Cypass supraciliary microstent (Alcon.com, Alcon Laboratories, a division of Novartis, Basel, Switzerland), the device featured in the COMPASS Trial, in terms of ECD loss rate and an enhanced risk of future cornea decompensation. At the time of this review, this device has been withdrawn from the market.

There are no current published RCT data on health‐related quality of life outcomes or visual field progression in people receiving supraciliary microstent surgery to achieve IOP‐lowering drop reduction.

Overall completeness and applicability of evidence

This review has shown that RCT evidence exists to assess the efficacy and safety of supraciliary microstent surgery for the treatment of OAG. The COMPASS trial has provided important data to support the primary outcome of this review and also several IOP‐associated secondary outcomes, importantly also including safety data (COMPASS TrialLass 2019; Reiss 2019). However, the COMPASS Trial only addresses one of the four subgroups of glaucoma interventions that the scope of this review set out to compare.

Although 60‐month safety and effectiveness data has been published, this study extension was not powered to allow statistical analysis beyond description to be presented. Twenty per cent of case data at 60 months required retrospective collection, the observations were unmasked, and only 253 of the original 505 cohort of cases completed the entire study extension period, raising additional concerns over selection bias (44% declined enrolment without a reason). At 60 months, endothelial cell density data were collected on only 163 participants from the 355 phacoemulsification + microstent cases that completed the initial 24‐month COMPASS Trial.

The results of another RCT, NCT01461278, are awaited, featuring an alternative supraciliary microstent.

Quality of the evidence

Although only one RCT exists (comprising 505 enrolled participants) relating to one of the five glaucoma intervention types sought in this review, the evidence presented was assessed to be of moderate‐ to high‐certainty. While the COMPASS Trial is well designed, this study also acknowledges the presence of unmasked investigators in the follow‐up period as a limitation, potentially introducing performance bias.

Potential biases in the review process

This review was conducted in line with the methods outlined by Cochrane. To ensure a high level of completeness in the search of electronic databases and trial registries, an Information Specialist was used. Selection of studies meeting the review inclusion criteria was performed independently by two of the review authors. Our review method ensured that only data from RCTs were included in this review.

Agreements and disagreements with other studies or reviews

We found no other systematic reviews to form a comparison.

original image

Figuras y tablas -
Figure 1

Risk of bias graph: review authors' judgements about each risk of bias item presented as percentages across all included studies.

Figuras y tablas -
Figure 2

Risk of bias graph: review authors' judgements about each risk of bias item presented as percentages across all included studies.

Risk of bias summary: review authors' judgements about each risk of bias item for each included study.

Figuras y tablas -
Figure 3

Risk of bias summary: review authors' judgements about each risk of bias item for each included study.

Comparison 1: Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone, Outcome 1: Proportion of participants medication‐free at 24 months

Figuras y tablas -
Analysis 1.1

Comparison 1: Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone, Outcome 1: Proportion of participants medication‐free at 24 months

Comparison 1: Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone, Outcome 2: Mean change in unmedicated IOP at 24 months

Figuras y tablas -
Analysis 1.2

Comparison 1: Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone, Outcome 2: Mean change in unmedicated IOP at 24 months

Comparison 1: Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone, Outcome 3: Mean change in number of IOP‐lowering medications taken per day

Figuras y tablas -
Analysis 1.3

Comparison 1: Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone, Outcome 3: Mean change in number of IOP‐lowering medications taken per day

Summary of findings 1. Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone for open‐angle glaucoma, at 24 months

Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone for open‐angle glaucoma

Patient or population: people with open‐angle glaucoma
Setting: hospital or outpatient clinic
Intervention: phacoemulsification + supraciliary microstent surgery
Comparison: phacoemulsification alone

Outcomes

Anticipated absolute effects* (95% CI)

Relative effect
(95% CI)

№ of participants
(studies)

Certainty of the evidence
(GRADE)

Comments

Risk with phacoemulsification alone

Risk with phacoemulsification + supraciliary microstent surgery

Proportion of participants who were medication‐free (not using eye‐drops)
at 24 months (medium‐term)

Study population

RR 1.27
(1.09 to 1.49)

448
(1 RCT)

⊕⊕⊕⊝
MODERATE1

 

595 per 1000 (500 to 685)

849 per 1000 (806 to 886)

Mean change in unmedicated IOP
24 months (medium‐term)

Study population

MD ‐2.0 mmHg  (‐2.85 to ‐1.15)

448
(1 RCT)

⊕⊕⊕⊕
HIGH2

 

The mean change (reduction) in IOP in the control group at 24 months was 5.4 (SD 3.9) mmHg

The mean change (reduction) in IOP in the intervention group at 24 months was 7.4 (SD 4.4) mmHg

Mean change in daily IOP‐lowering medications
at 24 months (medium‐term)

Study population

MD ‐0.5 medications (‐0.68 to ‐0.32)

448
(1 RCT)

⊕⊕⊕⊝
MODERATE1

 

The mean reduction in number of IOP‐lowering drops was 0.7 medications

The mean reduction in number of IOP‐lowering drops was 1.2 medications

Proportion of participants who required further glaucoma surgery
at 24 months (medium‐term)

Study population

RR 0.26 (0.07 to 1.04)

505
(1 RCT)

⊕⊕⊝⊝
LOW3

 

31 per 1,000

8 per 1,000

(2 to 32)

 

 

Mean change in health‐related quality of life

The included study did not report this outcome.

Rate of visual field progression or proportion of participants whose field loss progressed 

The included study did not report this outcome.
 

Proportion of participants experiencing postoperative complications
over 24 months (medium‐term)

Study population

RR 1.1 (0.8 to 1.4)

505
(1 RCT)

⊕⊕⊕⊕
MODERATE4

Five year post‐approval data regarding the safety of the Cypass supraciliary microstent showed increased endothelial cell loss, associated with the position of the microstent in the anterior chamber.

360 per 1,000

390 per 1,000

*The risk in the intervention group (and its 95% confidence interval) is based on the assumed risk in the comparison group and the relative effect of the intervention (and its 95% CI).

CI: confidence interval; IOP: intraocular pressure; MD: mean differenceOR: odds ratio; RR: risk ratio; SD: standard deviation

GRADE Working Group grades of evidence
High‐certainty: we are very confident that the true effect lies close to that of the estimate of the effect.
Moderate‐certainty: we are moderately confident in the effect estimate: The true effect is likely to be close to the estimate of the effect, but there is a possibility that it is substantially different.
Low‐certainty: our confidence in the effect estimate is limited: The true effect may be substantially different from the estimate of the effect.
Very low‐certainty: we have very little confidence in the effect estimate: The true effect is likely to be substantially different from the estimate of effect.

1Downgraded one level for study limitations: although participants were masked to their treatment group throughout the study period, as were IOP reading technicians, surgeons/Investigators were not.

2Not downgraded for study limitations as IOP assessment was masked.

3Downgraded two levels: one for imprecision ‐ confidence intervals included 1, no effect, and one level for risk of bias.

4Downgraded one level for imprecision: confidence intervals included 1, no effect.

Figuras y tablas -
Summary of findings 1. Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone for open‐angle glaucoma, at 24 months
Table 1. Ocular adverse events after phacoemulsification + supraciliary microstent surgery versus phacoemulsification, at 60 months

 

Intervention   
 

Phacoemulsification surgery
 

Phacoemulsification surgery + microstent
 

%

n

%

n

Outcomes

 

 

 

 

BCVA loss of > 3 lines or more compared to best BCVA reported in COMPASS study

0

0

0.9

2

Retinal detachment

1.5

1

0

0

Treatment of elevated intraocular pressure not satisfactorily managed with ocular hypotensive medication

1.5

1

0.5

1

Macular oedema

1.5

1

1.4

3

Other maculopathies

1.5

1

1.4

3

Corneal oedema

0

0

1.4

3

Events requiring unplanned surgical intervention

1.5

1

0.9

2

Results from the COMPASS XT publication. Phacoemulsification n = 200; phacoemulsification + microstent group n = 53, completing 60 months

Unmasked observational data. Similar baseline characteristics between groups.

Approximately 20% of the 60‐month data from the 253 participants cases completing the COMPASS XT study was obtained retrospectively.

Figuras y tablas -
Table 1. Ocular adverse events after phacoemulsification + supraciliary microstent surgery versus phacoemulsification, at 60 months
Table 2. Corneal endothelial cell loss after phacoemulsification + supraciliary microstent or phacoemulsification alone

Mean endothelial cell density (cells/mm2)        

 

Baseline 

At 60 months 

mean

Lower CI

Upper CI

n

mean

Lower CI

Upper CI

n

Phacoemulsification

2434.5

2356.5

2512.4

67

2189.1

2069

2309.2

40

Phacoemulsification + supraciliary microstent

2432.6

2382.8

2482.4

214

1931.2

1851.2

2011.2

163

 

Change from baseline (cells/mm2)     

mean

Lower CI

Upper CI

n

Phacoemulsification

‐249.6

‐341

‐158.2

40

Phacoemulsification + supraciliary microstent

‐507.6

‐581.7

‐433.6

163

 

Proportion of eyes with > 30% reduction in endothelial cell density from baseline at 60 months    

%

n

%

n

%

n

Number of retention rings visible on gonioscopy

< 1 microstent retention ring

1 microstent retention ring

≥ 2 microstent retention rings

Phacoemulsification + supraciliary microstent

20.6

13/63

21.9

16/73

57.7

15/26

Phacoemulsification alone

10.0

4/40

Results from the COMPASS trial safety extension publication. Extended interval specular microscopy performed subsequent to enrolment into the COMPASS extension trial (COMPASS XT)

Unmasked observational data

Figuras y tablas -
Table 2. Corneal endothelial cell loss after phacoemulsification + supraciliary microstent or phacoemulsification alone
Comparison 1. Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone

Outcome or subgroup title

No. of studies

No. of participants

Statistical method

Effect size

1.1 Proportion of participants medication‐free at 24 months Show forest plot

1

448

Risk Ratio (M‐H, Fixed, 95% CI)

1.27 [1.09, 1.49]

1.2 Mean change in unmedicated IOP at 24 months Show forest plot

1

448

Mean Difference (IV, Fixed, 95% CI)

‐2.00 [‐2.85, ‐1.15]

1.3 Mean change in number of IOP‐lowering medications taken per day Show forest plot

1

448

Mean Difference (IV, Fixed, 95% CI)

‐0.50 [‐0.68, ‐0.32]

Figuras y tablas -
Comparison 1. Phacoemulsification + supraciliary microstent surgery versus phacoemulsification alone